5.7 Adjusting the Temperature

1. From the Main Menu, select SENSOR SETUP and confirm.

2. Highlight the appropriate sensor if more than one sensor is attached and

confirm.

3. Select CALIBRATE and confirm.

4. Select PROCESS TEMP and confirm.

5. Press ENTER when Stable, TEMP: XX.X is displayed. confirm to continue.

6. Adjust the Reading XX.X °C with the keypad and confirm.

7. CAL COMPLETE, OFFSET: X.X °C, confirm to continue.

8. MOVE PROBE TO PROCESS is displayed. Confirm.

5.8 Calibrating two sensors simultaneously

1. Start by calibrating the first sensor and when you arrive at the point at which

you are prompted to "PRESS ENTER WHEN STABLE".

2. Press the BACK key.

3. Select EXIT and press accept. The instrument returns to the display of the

measurements. The measured value for the sensor to be calibrated starts to
flash.

4. Start the calibration of the other sensor and when you again arrive at the point

at which you are prompted to "PRESS ENTER WHEN STABLE".

5. Press the BACK key.

6. Select EXIT and press accept. The instrument returns to the display of the

measurements. The measured values for both sensors start to flash.

7. To return to the calibration menu for the individual sensors, press the Menu

key, select SENSOR SETUP and accept. Choose the required sensor and
accept.

When calibration is complete, accept.