Section 7 faults, causes, rectification, 1 error messages, 2 warnings – Hach-Lange 3798-S sc User Manual

Page 23: 1 error messages 7.2 warnings, 1 "error messages, 2 "warnings

Advertising
background image

21

Section 7

Faults, causes, rectification

7.1 Error messages

Possible sensor errors are displayed by the controller.

7.2 Warnings

Possible warning messages are displayed by the controller.

Table 1: Error messages

Error displayed

Cause

Rectification

*****

No communication with the
controller

Check the connection to the controller
Check the cable to the controller

SENSOR MISSING
FFFFFFFFFFFFF

No communication with the
controller

Check the connection to the controller
Check the cable to the controller

TEMP TOO LOW

Measured temperature < –5 °C

Ensure that the
medium temperature is > –5 °C.

TEMP TOO HIGH

Measured temperature >
+100 °C

Ensure that the
medium temperature is < +100 °C.

COND TOO LOW

Conductivity < 100

µS

/

cm

Ensure that the
conductivity is > 100

µS

/

cm

.

COND TOO HIGH

Conductivity > 500

mS

/

cm

Ensure that the
conductivity is < 500

mS

/

cm

.

RES. TOO LOW

Resistance < 2

Ω

Please contact service.

RES. TOO HIGH

Resistance > 10 k

Ω

Please contact service.

Table 2: Warnings

Error displayed

Cause

Rectification

CAL TOO OLD

The last calibration was more than
180 days ago.

Calibrate the sensor

HUMIDITY BAG

The desiccant bag is more than
1000 days old.

Please contact service.

Advertising