Technical data

1.1 Technical data for the 3798-S sc conductivity sensor

Specifications are subject to change without notice.

Materials

Stainless steel metal housing, PEEK

Enclosure rating

IP 68; stainless steel metal housing

Storage temperature
Sensor and controller

–20 °C ... 60 °C; 95 % relative humidity, non-condensing

Cell constant

K = 2.35 cm

–1

Measuring range conductivity

250

µS

/

cm

... 2.5

S

/

cm

Temp. measuring range

–5 °C ... 50 °C

Sensor operating temperature

–20 °C ... 50 °C

Conductivity response time

< 2 s; T90

Temp. response time

< 2 min; T90

Conductivity measuring accuracy

± 1 % of the measured value displayed or ± 0.004

mS

/

cm

Temp. measuring accuracy

± 0.2 °C

Reproducibility

< 0.2 %

Sensitivity

± 0.5 % of the end value of the measuring range

Sensor power

< 7 W

Calibration

Zero value calibration in air.
Fixed value calibration with defined resistance or with standard solution

Max. immersion depth / pressure
for the sensor

20 m / 2 bar

Maximum flow speed

4 m/s

Sensor interface

MODBUS

Sensor cable

10 m, hard wired, polyurethane

Sensor weight

< 1 kg

Sensor dimensions (Ш Ч L)

43 × 370 mm

Fastening

• Immersed pipe
• Chain