11

6.

If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.

7.

Push

EXIT

until the meter returns to the measurement mode.

Change check standard options

1.

Make sure an electrode is connected to the meter.

2.

Push the

METER OPTIONS

key and select (ISENa381) Settings.

3.

Select Modify Current Settings.

4.

Select Calibration Options.

5.

Update the settings. Refer to

Table 5

.

Table 4 Custom buffer sets

Buffer set values

Options

Description

Std1
Std2
Std3
Std4
Std5

10.0 mg/L

Pre-set temperature compensated standard values.

100.0 mg/L

1000.0 mg/L

2299.0 mg/L (0.1 M NaCl)

Custom

Custom standard value. Custom standard values are not
temperature compensated.

No standard

Standard is undefined when this option is selected.

Table 5 Check standard options settings

Settings

Options

Description

Standard

10.0 mg/L
100.0 mg/L (default)
1000.0 mg/L
0.1 M NaCl
Custom

Value for check standard.
Standard value will appear on Check
Standards Options screen.

Check Standard reminder

Reminder: On or Off (default)
Allow Defer:Yes

Meter automatically displays check
standard screen if reminder is on.
Measurement of check standard can be
delayed if set to Yes.

Acceptance criteria

Acceptance Limits: 1% to 20%
Cal Expires on Failure:Yes or No

Values that check standard must fall within
in order to pass.
Calibration expires if check standard fails.