Refer to, Troubleshooting – Hach-Lange HQD - Sodium (Na+) Ion Selective Electrode ISENa User Manual

Page 12

Advertising
background image

12

DOC272.53.80027

6.

If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.

7.

Push

EXIT

until the meter returns to the measurement mode.

Standard units

mg/L (default)
µg/L
g/L
g/kg
mol/L
mmol/L
mol/kg
%
ppm
ppb

Select the preferred unit for ISE check
standard.

Standard value
(for Custom standard option)

Enter standard value using up/down
arrow keys

Value for the custom check standard.

Table 5 Check standard options settings (continued)

Settings

Options

Description

Troubleshooting

Message or symptom

Possible cause

Action

Probe not supported

Software not updated

Update the HQd software to the newest revision at

www.hach.com/SoftwareDownloads.

Refer to the HQd

Series meter manual.

HQd meter does not support
IntelliCAL probe

Contact a Technical Support Representative.

Connect a probe or
probe requires service

Electrode not connected properly

Disconnect, then connect the electrode. Tighten the
locking nut.

Software not updated

Update the HQd software to the newest revision at

www.hach.com/SoftwareDownloads

. Refer to the HQd

Series meter manual.

Large number of methods stored on
electrode.

Continue to let electrode connect. Do not disconnect
electrode.

Damaged electrode

Verify connectivity with another electrode or meter to
confirm isolated issue with probe. Contact a Technical
Representative.

mV reading is the same
for all solutions

Soaker bottle not removed

Remove the soaker bottle.

Electrical issue

Contact a Technical Support Representative

Standard not recognized
error

Soaker bottle not removed

Remove soaker bottle.

Incorrect or contaminated buffer
solution

Use fresh standard solution as specified in the method

Contaminated filling solution

Replace the filling solution.

Slow response time

Tape not removed from the
filling-hole

Remove the tape.

Contaminated glass sensor

Clean the electrode.

Contaminated filling solution

Drain and replace the filling solution with fresh solution.

filling-hole cap closed

Open filling-hole cap while in use.

Low sample temperature or
temperature difference between
samples

Check the sample temperature. The lower the temperature
or the greater the difference of temperatures between
samples, traditionally the longer the response time.