Table 2 measurement options settings – Hach-Lange HQD - Sodium (Na+) Ion Selective Electrode ISENa User Manual

Page 9

Advertising
background image

DOC272.53.80027

9

Change measurement options
Methods are groups of default or user defined settings relevant to specific applications. If
the meter is set to the default method and the Modify Current Settings option is chosen, a
prompt will display to name the method after changes are entered. The settings are saved
with this name to distinguish them from the default method settings, which cannot be
changed. A saved method can be used instead of repeatedly adjusting individual settings.

Changes made to a user defined method are automatically saved with the existing name.
Multiple methods can be saved for the same probe.

1.

Make sure an electrode is connected to the meter.

2.

Push the

METER OPTIONS

key and select (ISENa381) Settings.

3.

Select Modify Current Settings.

4.

Select Measurement Options.

5.

Update the settings. Refer to

Table 2

.

Check standard options

Standard

Check standard reminder

Acceptance criteria

Table 1Parameter-specific settings

Setting

Options

Table 2 Measurement options settings

Setting

Options

Description

Units

mg/L (default)
µg/L
g/L
g/kg
mol/L
mmol/L
mol/kg
%
ppm
ppb

Select the preferred unit for ISE measurements.
Note: mV units are shown when the detailed display is
selected.

Significant digits

2
3 (default)
4

Significant digits are the digits of a number that are known
with certainty or have meaning that contributes to accuracy.

Auto stabilization

On (default)
Off

The default stability drift rate is 1.000 mV/min. Lower stability
criteria will require longer stabilization times, but the
measurement will be more precise. Higher stability criteria will
require shorter stabilization times, but the measurements may
be less precise.

Stability criteria

0.1 - 9.9 mV/min

Selectable

Measurement limits

Lower limit (default: 0.023 mg/L)
Upper limit (default: 14.00 mg/L)

The measurement limits can be set to match the acceptable
values for the sample. When the measurement is above the
upper limit setting or below the lower limit setting, the meter
will show an “Out of limits” message. This message can alert
the user to a potential problem with process conditions.

Advertising