Table 1 parameter-specific settings, Isa rinse solution preparation on, Isa rinse solution preparation – Hach-Lange HQD - Sodium (Na+) Ion Selective Electrode ISENa User Manual

Page 8: Advanced operations

Advertising
background image

8

DOC272.53.80027

Rinse or soak for 1 minute in 25 mL of 100 mg/L sodium standard (containing Sodium
ISA).

Calibrate and test the electrode again following measurement procedures.

Fill the electrode

1.

Open the filling-hole cap.

2.

Remove the cap from the tip of the filling solution bottle.

3.

Insert the dispensing tip into the filling-hole and position downward.

4.

Slowly squeeze the filling solution container and fill the electrode.

Note: Fill the electrode completely for optimum performance.

5.

Close the filling-hole cap until use and save the filling solution container and cap for
later use.

Note: If the dispensing tip becomes clogged, remove dispensing tip and soak in warm water. Dry
thoroughly and assemble.

Storage solution preparation
To prepare, add one Sodium Ionic Strength Adjustor (ISA) powder pillow (0.4 g) to 25 mL
of 100 mg/L Na

+

standard solution.

ISA rinse solution preparation
To prepare, add one Sodium Ionic Strength Adjustor (ISA) powder pillow (0.4 g) to every
25 mL of deionized water. Add prepared solution to rinse bottle and use as needed.

Advanced operations

Parameter-specific settings can be changed through the Full Access Options menu.
Details about menu navigation, available options and how to change them are given in the
screens, tables and procedures throughout this section.

The settings that can be changed are shown in

Table 1

.

Table 1Parameter-specific settings

Setting

Options

Measurement options

Units

Significant digits

Auto stabilization

Stability criteria

Upper and lower range limits

Calibration options

Standard set

Calibration units

Minimum calibration points

Slope limit

Calibration reminder

Advertising