Option

Function

Description

CLr ALL
CAL

Clear calibration
points

Erases all the calibration points. Make sure to calibrate the meter.

1

int

Data log interval

Keeps up to 999 data points in intervals from 1 to 1999 seconds (default =
10) in the data log.

year

Year

Use the number keys to enter the correct year.

date

Date format

Sets the date to mm/dd/yy or dd/mm/yy format.

m/d date

Month and day

Use the number keys to set the correct month and day.

time

Time

Sets the correct time in a 24-hour format. Use the number keys to set the
correct time.

oFF

Automatic
shutdown

Sets the shutdown parameters from 1 minute to 000 minutes (continuous
power). The meter beeps 1 minute before shutdown. Make sure to power
cycle for the automatic shutdown to occur. Automatic shutdown is disabled
during: data logging, Bluetooth

transfers and when connected to the USB

port.

Snd

Sound options

Sets sound alerts to on or off. Three sounds tell the user about different
functions:

One beep: Stabilization occurred with the stabilization lock on.
Two beeps: An error occurred. The error code shows on the display.

Refer to

Troubleshooting

on page 17.

Three beeps: Measurement stability in calibration mode, regardless of

the stabilization lock setting.

ºC ºF

Temperature
display

Sets the temperature unit from ºC to ºF.

READY

Stabilization lock

Sets the stabilization lock to on or off:

• The "READY" icon shows when an endpoint occurs and the stabilization

lock is set to on.

• After stabilization, the display locks the value and the meter ignores slight

measurement changes in pH, conductivity or TDS.

• Sets the stabilization lock to off during titrations or slight change

detection. The display automatically unlocks after a significant
measurement change is found.

1

This does not erase calibration data for an ISFET probe.

4. Push ENTER to keep the changes and go back to setup mode.

Note: To exit and not keep the changes, push CANCEL.

Operation

Configure the Bluetooth

wireless connection

Transfer data and control several instrument functions from a remote location to a PC with the
Bluetooth

wireless connection.

Note: The arrow icon shows on the display when the meter is in communication with the SmartLogger II software
on a PC. The arrow icon flashes when data is transferred. Refer to the SmartLogger II documentation to setup the
wireless connection to a PC.

1. Set the power to on. The meter looks for the Bluetooth

connection with a PC.

2. On the PC, select the option to find or add a Bluetooth

device. Next, the PC prompts if the user

wants to pair with a found device.

3. When the PC prompts for a pass key or PIN, enter the PIN (default = 1234). The PC shows if the

pair is successful and the meter shows the Bluetooth

icon.