Error codes – Hach-Lange H-Series Meters Basic User Manual User Manual

Page 19

Advertising
background image

Error codes

Table 2

shows the codes that can occur for various reasons. Error codes show instrument

malfunction or user error.

Table 2 Error code descriptions

Error
Code

Description

Solution

E02

The ISFET pH probe is damaged.

Replace the probe. Contact technical support for probe
replacement information.

E03

Clean the probe.

Clean the probe. If the error continues, replace the probe.

E04

Glass pH probe slope error. The
slope is smaller than 85% or larger
than 102% of 59.16 mV per pH unit.

Clean the probe. If the error continues, replace the probe.

E06

ISFET pH probe slope error.

Clean the probe. Soak in 40 ºC (113 ºF) pH 7.00 buffer for
2 minutes. Calibrate again. If the error continues, replace the
probe.

E07

Clean the probe.

Clean the probe. Make sure that no air bubbles on the sensor
surface or foreign objects or materials on the sensor.

E08

Too long to calibrate. Signal is not
stable during calibration.

Clean the probe. Disconnect stirrers and other AC power
sources. Make sure that the calibration solution temperature
is constant. Replace the probe if the error continues.

E13

The ISFET pH probe temperature
sensor is damaged.

Replace the probe. Contact technical support for probe
replacement information.

E14

The battery is extremely low.

Replace the batteries immediately. Damage to the accuracy
and function of the meter could have occurred.

E15

Replace the battery immediately.
The accuracy and function of the
meter could be compromised.

Contact technical support for service information.

E20

The conductivity temperature
sensor is damaged.

Replace the probe. Contact technical support for probe
replacement information.

E25

The slope is smaller than 60% or
larger than 140% of the nominal.

Replace DO probe membrane and fill solution. Replace the
probe if the error continues. Nominal = 0.1822 ppm/mV.

E26

The mV reading is more than
±10 mV from nominal.

Replace DO probe membrane and fill solution. Replace the
probe if the error continues. Nominal = 0.0 mV at 0%
saturation; 45mV at 100%.

E27

DO probe temperature error

Attach the temperature sensor to the 3.5 mm phono jack. DO
readings are highly temperature dependent and a
temperature sensor must be attached.

E28

Barometric pressure sensor error

Contact technical support for repair information.

E30

ISE electrode calibration error

The slopes are not the same sign or are not within 25% of
each other. Calibrate again in the correct solutions.

E40

Unrecognized host command

Command from a host PC is not found. Use only valid
commands.

E42

Invalid input

The value entered during the setup is invalid. Enter a
different value.

E44

No probe is installed

No probe is installed for the applicable parameter. Set the
meter power to off. Install the correct probe. Set the meter
power to on.