Startup, Turn the meter on and off, Operation – Hach-Lange SENSION+, Portable Meter, LPG2500 User Manual

Page 10: Setup for ph, Setup for dissolved oxygen, Polarize the electrode, Change the settings, Setup for conductivity

Advertising
background image

Startup

Turn the meter on and off

N O T I C E

Make sure that the probe is connected to the meter before the meter is turned on.

Push and hold to turn on or turn off the meter. If the meter does not
turn on, make sure that the batteries are properly installed.

Note: The meter is set to automatically power off after 5 minutes of inactivity to
maximize the battery life.

Operation

W A R N I N G

Chemical exposure hazard. Obey laboratory safety procedures and wear all of
the personal protective equipment appropriate to the chemicals that are handled.
Refer to the current material safety data sheets (MSDS) for safety protocols.

Setup for pH

The meter is ready for pH or ORP calibration without additional setup.

Setup for dissolved oxygen

Before calibration or measurement for dissolved oxygen, the probe must
be polarized and the settings for atmospheric pressure and salinity must
be entered.

Polarize the electrode

If the probe or batteries are disconnected, connect the probe or install
the batteries and wait for polarization:

Disconnect time

Polarization time

< 5 minutes

10 minutes

5 to 15 minutes

45 minutes

> 15 minutes

6 hours

Change the settings

The settings for dissolved oxygen measurements can be changed when
a dissolved oxygen probe is used. Push

to view the current settings.

1. Push

. The first setting is shown.

2. Quickly use the arrow keys to change the value. Wait for the next

setting to be shown and change any additional values:

Option

Description

Atmospheric pressure in mbar (default: 1013 mbar)

SAL

Salinity in g/L NaCl (default: 0 g/L)
Aut: automatic (only in model MM156)

SEn

Select the DO sensor model: 5120 or 5130

Note: The salinity value can be found from a conductivity measurement. Refer
to

Table 3

on page 17.

Setup for conductivity

The settings for conductivity measurements can be changed when a
conductivity probe is used. Push

to view the current settings.

1. Push

. The first setting is shown.

2. Quickly use the arrow keys to change the value. Wait for the next

setting to be shown and change any additional values:

Option Description

CEL

Electrode type: platinum (default) or titanium.
Note: This option is not available on all meters.

CAL

The number of standards for conductivity calibration. Select one
standard (default). Select the standard nearest to the expected
measurement range.

tC

Temperature compensation: 0 to 9.99%/°C (default: 2%/°C)

tr

Reference temperature: 20 or 25 °C (default: 25 °C)

F

TDS calculation factor: 0.01 to 4.44 (default: 0.64)

10 English