Table 1 pH, ORP (mV) and temperature values

Temperature

pH

mV

°C

°F

0

32

2.01

4.01

7.12

9.52

10.30

10

50

2.01

4.00

7.06

9.38

10.17

245

20

68

2.00

4.00

7.02

9.26

10.06

228

25

77

2.00

4.01

7.00

9.21

10.01

220

30

86

2.00

4.01

6.99

9.16

9.96

212

40

104

2.00

4.03

6.97

9.06

9.88

195

50

122

2.00

4.06

6.97

8.99

9.82

178

60

140

2.00

4.10

6.98

8.93

9.76

160

70

158

2.01

4.16

7.00

8.88

80

176

2.01

4.22

7.04

8.83

90

194

2.01

4.30

7.09

8.79

Conductivity standard solutions
Refer to

Table 2

for conductivity values of standard solutions at varying

temperatures.

Table 2 Conductivity and temperature values

Temperature

Conductivity (EC)

°C

°F

µS/cm

µS/cm

mS/cm

mS/cm

15.0

59

119

1147

10.48

92.5

16.0

60.8

122

1173

10.72

94.4

17.0

62.6

125

1199

10.95

96.3

18.0

64.4

127

1225

11.19

98.2

19.0

66.2

130

1251

11.43

100.1

Table 2 Conductivity and temperature values (continued)

Temperature

Conductivity (EC)

°C

°F

µS/cm

µS/cm

mS/cm

mS/cm

20.0

68

133

1278

11.67

102.1

21.0

69.8

136

1305

11.91

104.0

22.0

71.6

139

1332

12.15

105.4

23.0

73.4

142

1359

12.39

107.9

24.0

75.2

145

1386

12.64

109.8

25.0

77

147

1413

12.88

111.8

26.0

78.8

150

1440

13.13

113.8

27.0

80.6

153

1467

13.37

115.7

28.0

82.4

156

1494

13.62

29.0

84.2

159

1522

13.87

30.0

86

162

1549

14.12

31.0

87.8

165

1581

14.37

32.0

89.6

168

1609

14.62

33.0

91.4

171

1638

14.88

34.0

93.2

174

1667

15.13

35.0

95

177

1696

15.39

16 English