Hach-Lange SENSION+, Portable Meter, LPG2500 User Manual

Page 14

Advertising
background image

Error/Warning

Description

Solution

Measurement out
of range.

• Insert the probe into the appropriate

standard and read again.

• Clean the probe and the membrane.
• Make sure that there are no air

bubbles in the membrane.

• Disconnect the probe and then

connect the probe again.

• Connect a different probe to verify if

problem is with probe or meter.

Battery is low.

Insert new batteries.

E1

Unstable reading
during a stability
measurement or
during calibration.

Make sure that the probe is properly
immersed in the sample.

E2

Probe current:
exceeded upper
limit during
measurement ≥
250 nA.

• Clean the probe and the membrane.
• Make sure that there are no air

bubbles in the membrane.

• Disconnect the probe and then

connect the probe again.

• Connect a different probe to verify if

problem is with probe or meter.

E2

pH probes
sensitivity out of
range (accepted
values 70–105%).

• Make sure that the correct probe is

connected to the meter.

• Connect a new probe.

Error/Warning

Description

Solution

E2

During calibration,
the difference
between the
constants gets >
30%.

• Insert the probe into the appropriate

standard and read again.

• Clean the probe and the membrane.
• Make sure that there are no air

bubbles in the membrane.

• Disconnect the probe and then

connect the probe again.

• Connect a different probe to verify if

problem is with probe or meter.

• Connect a new probe.

E3

During calibration,
the cell constant
gets < 0.05 cm

-1

.

• Insert the probe into the appropriate

standard and read again.

• Clean the probe and the membrane.
• Make sure that there are no air

bubbles in the membrane.

• Disconnect the probe and then

connect the probe again.

• Connect a different probe to verify if

problem is with probe or meter.

E3

Slope out of range
(accepted values ±
58 mV).

Connect a new probe.

E4

Buffer solution not
recognized.

• Clean the probe and the membrane.
• Make sure that there are no air

bubbles in the membrane.

• Disconnect the probe and then

connect the probe again.

• Connect a different probe to verify if

problem is with probe or meter.

• Make sure that the buffer used

matches the buffer specified in
configuration.

• Replace the buffer solution.
• Make sure of the temperature

specification in configuration.

14 English