Salinity values for dissolved oxygen meters – Hach-Lange SENSION+, Portable Meter, LPG2500 User Manual

Page 17

Advertising
background image

Salinity values for dissolved oxygen meters

Table 3

can be used to convert conductivity values to salinity for

dissolved oxygen meters.

Table 3 Conversion from conductivity to salinity

Conductivity (mS/cm)

Salinity (g/L NaCl)

1.9

1.0

3.7

2.0

7.3

4.0

10.9

6.0

17.8

10.1

25.8

15.1

33.6

20.2

41.2

25.4

48.9

30.6

56.3

35.8

62.2

40.0

69.4

45.3

75.7

50.0

English 17