Replacement parts and accessories, Standard solutions – Hach-Lange SENSION+, Portable Meter, LPG2500 User Manual

Page 15

Advertising
background image

Error/Warning

Description

Solution

E4

Salinity cannot be
calculated TC=0

Modify TC

E5

Same buffer
solutions.

• Clean the probe and the membrane.
• Make sure that there are no air

bubbles in the membrane.

• Disconnect the probe and then

connect the probe again.

• Connect a different probe to verify if

problem is with probe or meter.

• Make sure that the buffer used

matches the buffer specified in
configuration.

• Replace the buffer solution.
• Make sure of the temperature

specification in configuration.

E6

The calibration
solutions have
different
temperatures.

Make sure that the calibration solutions
have the same temperature.

ErA (only
MM156)

Salinity is set to
automatic and
measured value >
50 g/L.

Modify TC

Replacement parts and accessories

Note: Product and Article numbers may vary for some selling regions. Contact the
appropriate distributor or refer to the company website for contact information.

Replacement parts

Description

Item no.

Batteries, Alkaline AA

1938004

Carrying case

LZW9990.99

Accessories

Description

Item no.

Electrolyte for DO probe 51 20, 25 mL

LZW9811.99

Electrolyte for DO probe 51 30, 50 ml

2759126

Tube for DO probe calibration

LZW5123.99

Service kit for 51 30 DO probe (contains 2 membrane
modules and DO filling solution)

5196800

Replacement DO membrane for DO probe 51 20

LZW5125.99

147 μS/cm standard solution (at 25 °C, 77 °F), 125 mL
bottle

LZW9701.99

1413 μS/cm standard solution (at 25 °C, 77 °F), 125 mL
bottle

LZW9711.99

12.88 mS/cm standard solution (at 25 °C, 77 °F), 125 mL
bottle

LZW9721.99

Set of calibration tubes, 147 μS/cm, 1413 μS/cm and
12.88 mS/cm

LZW9138.99

220 mV standard solution, 125 mL

LZW9402.99

Buffer solution pH 4.01, 125 mL

LZW9460.99

Buffer solution pH 7.00, 125 mL

LZW9461.97

Buffer solution pH 10.01, 125 mL

LZW9470.99

Electrolyte solution, KCI 3M, 50 mL

LZW9509.99

Calibration tubes pH 4.01, pH 7.00 and pH 10.01

LZW9137.97

Tube for ORP standard calibration 220 mV

LZW9136.99

Work protector for electrodes 50 51 T and 50 52 T

LZW9162.99

Standard solutions

Technical buffer solutions (DIN 19267)
Refer to

Table 1

pH and ORP (mV) values of specific buffer sets at

varying temperatures.

English 15

Advertising