Hach-Lange FP 360 sc Basic User Manual User Manual

Page 12

Advertising
background image

English 12

SENSOR SETUP

SELECT SENSOR (if there is more than one sensor)

CALIBRATE

SET OUTMODE

Behavior of the outputs during calibration and zero
point adjustment

HOLD

ACTIVE

SET TRANSFER

SENSOR MEASURE

Current, uncorrected measurement value

CONFIGURE

FACTOR: 0.1 to 100
OFFSET: –1000 to +1000

FACTOR

OFFSET

2 POINT

3 POINT

4 POINT

5 POINT

FACTOR

Is shown when FACTOR has been selected under
CONFIGURE. Refer to

Calibration on page 13

for

detailed information.

OFFSET

Is shown when OFFSET has been selected under
CONFIGURE. Refer to

Calibration on page 13

for

detailed information.

2-POINT

Is shown when 2 POINT has been selected under
CONFIGURE. Refer to

Calibration on page 13

for

detailed information.

3-POINT

Is shown when 3 POINT has been selected under
CONFIGURE. Refer to

Calibration on page 13

for

detailed information.

4-POINT

Is shown when 4 POINT has been selected under
CONFIGURE. Refer to

Calibration on page 13

for

detailed information.

CALIBRATE

5-POINT

Is shown when 5 POINT has been selected under
CONFIGURE. Refer to

Calibration on page 13

for

detailed information.

SET CAL DEFLT

Security query, reset to (FACTOR=1, OFFSET=0)

CONFIGURE

EDIT NAME

Name can include up to 16 characters
DEFAULT CONFIG: sensor serial number

SET PARAMETER

PAH: Measurement value related to PAH calibration
standard
OIL: Measurement value related to oil calibration
standard
DEFAULT CONFIG: PAH

SELECT UNITS

ppb, ppm, µg/L, mg/L,
DEFAULT CONFIG: ppb

AVERAGE

1 to 300 s,
DEFAULT CONFIG: 3 s

LOG SETUP

5 s, 30 s, 1, 2, 3, 4, 5, 6, 10, 15, 30 min,
DEFAULT CONFIG: 10 min

GAIN VALUE

Range 0 to 500 ppb: AUTO, 0.01 to 50, 0.01 to 500,
Range 0 to 5000 ppb: AUTO, 0.01 to 500, 0.01 to
5000
DEFAULT CONFIG: AUTO

SET DEFAULTS

Security query,
reset to default configuration for all menu options
listed above.

SENSOR SETUP (Continued)

SELECT SENSOR (if there is more than one sensor)

Advertising