Calibration – Hach-Lange FP 360 sc Basic User Manual User Manual

Page 13

Advertising
background image

English 13

Calibration

Factory calibration

The calibration curve zero point and slope are preset. Retrospective
calibration of these basic settings is generally not required outside of the
inspection intervals.
Do regular zero point checks to make sure that impurities or faults are
being detected (refer to

Check the zero point on page 14

).

If the zero signal increases due to the measuring medium components
or the installation conditions, you can compensate for this influence via
an offset correction. Carry out a lab analysis of the sample to do this. If
there is no PAH/oil contamination in the measuring medium, enter the
measurement value shown by the device as the offset (refer to

Adjust

the zero point (OFFSET) on page 15

).

Process calibration/adjustment

The sensor is pre-calibrated with various concentrations of a special
calibration standard in ultra-pure water. These ideal measurement
conditions rarely occur in reality. The measurement values shown are
qualitative trend indicators if no adjustment is made to the on-site
measurement conditions.
If you require quantitatively correct measurement values, either a
contrast adjustment or a multi-point calibration needs to be carried out.
Both of these operations must be carried out on site using lab analysis
data. Basic prerequisites for quantitative measurements are precise
knowledge about the oil type in occurrence and constant measurement
conditions, e.g. in cool water in a heat exchanger. If several oils are
present in varying quantities, it is generally not possible to carry out a
quantitative measurement.
If the measurement conditions change, you must check the accuracy of
the results again through lab analyses and make any adjustments where
necessary.

DIAG/TEST

PROBE INFO

SENSOR NAME

Device name

EDIT NAME

SERIAL NUMBER

Device serial number

RANGE

0 to 500 or 0 to 5000

MODEL NUMBER

Item no. Sensor

CODE VERSION

Sensor software

COUNTER

OPERATING
HOURS

Operating hours counter

MAINTENANCE

Counter counting down days

BULB CHANGE

Counter counting down days

TEST/MAINT

SET OUTMODE

Equipment output behavior in the SERVICE menu

HOLD

ACTIVE

SET
TRANSFER

SIGNALS

LAMP CURR

Flash lamp intensity

DIAG / TEST

Zero point and slope check with external standards

READING
OFFSET

CUBE CAL

Password-protected access for the service

SENSOR SETUP (Continued)

SELECT SENSOR (if there is more than one sensor)