English 17

Troubleshooting

Error messages

Possible sensor errors displayed by the sc controller.

Warnings

Possible sensor warning messages displayed by the sc controller.

Replacement parts

Displayed errors

Definition

Resolution

SENSOR ERROR

Electronic defect

Call manufacturer customer
service

Displayed warnings Definition

Resolution

DIAG/TEST

Counter expired

Call manufacturer customer
service

BULB CHANGE

Counter expired

Call manufacturer customer
service

LAST CONFIGUR

Changed configuration
was not accepted

Send configuration again

TARGET VALUE

With multi-point
calibration, values not
entered in ascending
order

Enter calibration values in
ascending order

Designation

Quantity

Service life

Flash bulb

1

4 years

O-Rings

4

2 years