4 verification kit, Advanced operations – Hach-Lange DR2800 eco User Manual

Page 106

Advertising
background image

106

Advanced Operations

6.7.3.4 Verification kit

The Verification Kit (see

Section 9 on page 123

) is designed for

periodic monitoring of scattered light, photometric accuracy and the
wavelength accuracy of the spectrophotometers.

When results exceed allowable tolerances (given in the quality
control certificate to the test record), contact the manufacturer.

1. Press Verification Kit in the Optical Checks menu.

2. Press Nominal Values.

3. Press Edit.

An automatic menu guidance queries values (filters,
wavelength, nominal values and tolerances) given in the quality
control certificate, to the following specifications:

Stray Light

Photometrical accuracy

Wavelength accuracy

4. Press OK when all values are entered and the overview is

displayed.

5. Press Verification.

6. Insert the adapter A (

Figure 3 on page 15

) in cell

compartment #2.

Advertising