Advanced operations – Hach-Lange DR2800 eco User Manual

Page 107

Advertising
background image

107

Advanced Operations

7. Remove any cuvettes/cells from the cell compartment and

press Start.

8. Insert the different filter in the given order one after the other.

Press Next after inserting a filter.

After the last measurement the results are displayed.

9. Press PC & Printer icon to send the data to a USB memory

stick, PC or to a printer.

The files will be stored automatically as CSV file (Comma
Separated Value). The file name will be formatted as
“Verification.csv“.