Hach-Lange LINK2SC User Manual

Page 45

Advertising
background image

English 45

f.

Use JOB->LAB to transfer the job file from the job status
menu to the laboratory.
(SD memory card must be inserted in the sc controller if this is
being used.)

2.

Process the job with the photometer (refer to

section Work

sequence: photometer

).

(If the SD memory card is being used, it must be connected to the

photometer via a USB adapter.)
a.

Use the LINK2SC button on the toolbar to open the job list.

b.

Select a job.

c.

Select the parameter to be processed.

d.

Analyze the comparison sample.

e.

Assign the result to the job.

f.

Use SEND TO SCto return the processed job.
(If the SD memory card is being used, it must be connected to
the photometer via a USB adapter.)

3.

LAN

1

: When the default setting is active, the probe is automatically

corrected with the laboratory measurement data once the job has

been returned.
Note: LINK2SC>CONFIGURE>AUTO JOB can be used to
deactivate automatic correction. The probe must then be corrected
manually as per transfer via SD memory card.

or
SD memory card: The job must then be activated manually on the
controller. Proceed as follows:
a.

Insert the SD memory card in the sc controller.

b.

Select LINK2SC from the sc controller menu.

c.

Select JOBS FROM CARD from the LINK2SC menu. (Jobs
are loaded into the sc controller.)

d.

Select JOB LIST from the LINK2SC menu.

e.

Select the corresponding job from the list.

f.

Select ACTIVATE JOB. (The probe is corrected.)

Work sequence: NEW MEASUREMENT

4.

In order for Link2SC to be able to select the best correction type for

the application in question (refer to

section Introduction to matrix

correction

), each further sensor correction must be performed via

"NEW MEASUREMENT" and not via "CREATE JOB".
a.

Select LINK2SC from the sc controller menu.

b.

Select JOB LIST from the LINK2SC menu.

c.

Select the corresponding job from the list.

d.

Select NEW MEASUREMENT.

e.

Select the PARAMETER for the new measurement.

f.

Take a comparison sample for the laboratory measurement.

g.

Use JOB->LAB to transfer the job file from the job status
menu to the laboratory.
(SD memory card must be inserted in the sc controller if this is
being used.)

5.

Process the job with the photometer (refer to 2. a. to f.)

6.

LAN

1

: The probe is automatically corrected with the laboratory

measurement data once the job has been returned.
or
SD memory card: The job must then be activated manually on the
controller. (refer to 3. a. to f.)

1. Only valid for sc1000 controllers.