background image

English 48

4.

Perform the parameter measurement in line with the work

procedure.

5.

Press the LINK2SC button to assign the measurement value to the

job.

6.

Press OK to close the job.

7.

Press SEND TO SC to return the job to the sc controller.
a.

LAN: The probe is automatically corrected with the laboratory
measurement data once the job has been returned.

Note: Once the correction has been implemented successfully,
confirmation (job with green tick) is sent back to the photometer,
where it can be archived. If an implausible measurement value
means that the correction cannot be made, the job is returned with a
red cross.

Note: If the network connection is interrupted, an error message is
output. The job can also be saved on a connected SD memory card
and transferred to the controller that way.

b.

SD memory card: If the SD memory card is being used, it
must be connected to the photometer via a USB adapter.
The job must then be manually activated on the controller.