5.2 Opening the pipette (1–5 mL)

1. Remove ejector button and tip ejector as described above

(for 0.2–1.0 mL pipettes)

section 5.1 on page 19

.

2. Remove cover.

3. Remove cone and cylinder with the help of the tool by

pressing the two projections together. While pressing the
projections, push the tool towards the cone.

4. Take apart, clean and grease as described above (for

0.2–1.0 mL pipettes)

section 5.1 on page 19

.

5. Reassemble in reverse sequence.

Figure 5 Opening the pipette

1

Service tool to open the pipette