Calibration – Hach-Lange Piston pipette User Manual

Page 25

Advertising
background image

23

Calibration

2. Use light pressure to place a tip on the tip cone, turning

the tip slightly.

3. Pipette distilled water into previously weighed beaker at

least five times. Determine each weight in mg to the first
decimal place and compare this with the weight shown
above. If the weight is not within this range, recalibrate the
pipette.

AV_pipette.book Seite 23 Donnerstag, 8. Dezember 2005 3:56 15