Section 6 calibration, 1 checking the calibration – Hach-Lange Piston pipette User Manual

Page 24

Advertising
background image

22

Section 6

Calibration

The pipette is calibrated in the factory with distilled water at
22 °C/72 °F. No recalibration is needed for routine use. The
pipette can, however, be recalibrated for other temperatures
and solutions of different viscosities. An appropriately
sensitive analytical balance, a small beaker and distilled water
are needed to carry out the calibration.

Important Note: If either the handle or the piston is replaced,
the pipette must be recalibrated. If the O-rings or the tip cone
are replaced, the calibration should be checked.

6.1 Checking the calibration

1. Set the pipette volume as follows in

Table 1 Checking the

calibration

:

Table 1 Checking the calibration

Pipette

(µL)

Volume setting

permissible

weight range

(mg)

10

10

9.8–10.2

20 20

19.8–20.2

50

50

49.5–50.5

100 100

99.2–100.8

200

200

198.7–201.3

500 500

497.0–503.0

1000

1000

994.0–1006.0

2000 2000

1990.0–2010.0

200–1000

300

298–302

1000–5000 2000 1990–2010