Figure 4 Instrument dimensions

1 22 cm (10 in.) minimum clearance

Electrical installation

Connect an sc controller using Quick-connect fitting

The instrument can be connected to an sc controller using the keyed
quick-connect fitting (

Figure 5

). Extension cables are available. Retain

the connector cap to seal the connector opening in case the instrument
must be removed. Refer to the sc controller manual for more information.

Figure 5 Quick-connect fitting

Plumbing

Sample line considerations

Select a good, representative sampling point for optimum instrument
performance. The sample analyzed must be representative of the
conditon of the entire system.
To minimize erratic readings:

• Draw samples from locations that are sufficiently distant from points of

chemical additions to the process stream

• Make sure samples are sufficiently mixed
• Make sure all chemical reactions are complete

Connect the sample stream

Connect sample inlet and drain lines to the fittings on the body of the
instrument (

Figure 1

on page 6). Sample inlet and drain tubing are

supplied by the customer. Refer to

Specifications

on page 3 for tubing

requirements.
Install the sample line(s) into a larger process pipe to minimize
interference from air bubbles or pipeline bottom sediment. A sample line
going into the center of a process pipe is best.

8 English