WARNING:

Wear appropriate personal protective equipment when handling
parts that come into contact with potentially biohazardous
samples.

To decontaminate the FLEXMAP 3D

®

instrument:

1. Remove all specimens and all FLEXMAP 3D

®

reagents.

2. Use the software to run a sanitize command with the diluted (10% to 20%) household

bleach solution. Leave deionized water and household bleach solution diluted to 10% to
20% in water in the system.

3. Use the software to run two wash commands with deionized water.

4. Turn off the power switch and unplug the cord from the outlet.

5. Wash all exterior surfaces with a mild detergent followed by the diluted (10% to 20%)

household bleach solution.

6. Open the front doors of the system. Clean all accessible surfaces with mild detergent

followed by the diluted (10% to 20%) household bleach solution.

Installation of Instrument

Luminex handles the packing, shipping, unpacking, and installation of the Luminex

®

FLEXMAP 3D

®

system. Luminex recommends that users and laboratory personnel do not

uninstall, move, or install the system.

For site preparation and facilities requirements, see the section titled

Installation

.

FIGURE 7.

Installed FLEXMAP 3D

®

System

Luminex

®

FLEXMAP 3D

®

Hardware User Manual

8