Continued – Luminex FLEXMAP 3D User Manual User Manual

Page 33

Advertising
background image

Temperature control:

Maintains samples using the heater block at a constant temperature from 35°C to 60°C
(95°F to 131°F), +/- 1°C of setpoint.

Altitude:

Operation up to 2400 m (7874 ft.) above mean sea level

Automatic transfer of assay protocols and new reagent information into the system using a large capacity read/write DVD

Automatic sampling from a 96- or 384-well microtiter plate, beginning from any well position

Automatic real-time analysis

Analyzes multiple assay protocols per microtiter plate

Barcode reader entry of sample IDs

System initialization:

< 45 minutes (including laser warmup, required software procedures, and weekly
calibration)

System verification:

5 minutes

FLEXMAP 3D

®

warmup:

30 minutes. A system that remains inactive for at least four hours requires a warm-up to
restart the lasers. The system resets the four-hour internal clock after acquiring the sample,
running system calibrators, running system controls, or warming up the instrument.

Produces sound pressure levels below 85 dBA

Operating Shock:

The instrument will be able to operate while experiencing limited shock.

• Pulse amplitude: 1 m/s

2

maximum

• Pulse direction: along any axis defined by a rear corner of the instrument, in either

direction

• Reference IEC 60068-2-27 Basic Environmental Testing Procedures Part 2: Tests -Test

Ea and Guidance: Shock

Note: This specification is not meant to comprehensively describe all forms of shock the
instrument may be exposed to in use; it is only intended to serve as a measure of the
system's robustness.

Although the system shall be able to withstand a shock pulse as described above without
compromising reliable operation, extended exposure to shock whether or not it exceeds the
pulse defined above is not recommended and may result in instrument failure.

Operating Vibration:

The instrument will be able to operate while experiencing limited vibration.

• Direction shall be along any axis defined by a rear corner of the instrument.

• Sinusoidal, limited to frequency ranges in table below

Frequency Range Limit Type of Limit

0 to 10 Hz 0.35 mm ± 10%

(0.7 mm peak-to-peak) Displacement

10 Hz to 100 Hz 1 m/s

2

± 10%

peak Acceleration

Reference IEC 60068-2-6 Environmental Testing Procedures Part 2: Tests - Test Fc:
Vibration (Sinusoidal).

Note: Although the instrument shall be able to operate while experiencing vibration as
described above, extended exposure to vibration whether or not it exceeds these limits is
not recommended and may result in instrument failure.

TABLE .

Continued

For research use only. Not for use in diagnostic procedures.

Technical Overview

23

Advertising