Luminex FLEXMAP 3D User Manual User Manual

Page 43

Advertising
background image

9. Screw the syringe back into its housing.

10. Return the syringe arm to its original position. The bottom of the plunger fits into the

indentation in the syringe arm.

11. Hand-tighten the set screw on the syringe arm.

12. Plug in the power cord and turn on the FLEXMAP 3D instrument.

13. Use the software to run the prime command twice, watching for any leaks in the syringe

area.

14. Close the right front door.

FIGURE 16.

Syringe Assembly (Inside Right Door)

1 Glass Syringe Barrel

4 Syringe Plunger - Pull out as indicated in Step 6 and

replace as referenced in Step 9.

2 Syringe Arm (between dotted lines) - Press down as

indicated in Step 4 and up as referenced in Step 10.

5 Plunger Seal (contains O ring) - Replace as

indicated in Step 7.

3 Set Screw - Loosen as indicated in Step 4 and

tighten as referenced in Step 11

For research use only. Not for use in diagnostic procedures.

Maintenance and Cleaning

33