Bead detail irregularities continued – Luminex FLEXMAP 3D User Manual User Manual

Page 62

Advertising
background image

Symptom

Description

Possible Problem

Solution

xMAP microspheres do not
form a cohesive group and are
somewhat scattered.

There is air in the instrument.

Verify the sample probe height. Use the
software to run a prime command three
times, an alcohol flush command twice, then
a wash command three times with deionized
water.

xMAP microspheres are
widely scattered.

The sheath fluid is empty.

Make sure there is sheath fluid in the sheath
container. Use the software to run a prime
command until all air is pushed out of the
instrument.

TABLE 9.

Bead Detail Irregularities Continued

Luminex

®

FLEXMAP 3D

®

Hardware User Manual

52

For research use only. Not for use

in diagnostic procedures.

Advertising