Adjust the sample probe vertical height – Luminex 200 Installation Guide User Manual
Page 38

3. Unscrew the cap from the Cheminert fitting. See the Sample Probe and Cheminert Fitting
image below.
FIGURE 12.
Sample Probe and Cheminert Fitting
1
Cheminert Fitting
4
Adjustment Slide
2
Sample Probe Holder
5
Sample Probe
3
Thumbscrew
4. Insert the sample probe into the sample probe holder. Insert the Cheminert fitting into the
sample probe holder, turning it clockwise. Be careful that the threads are correctly
aligned. Hand tighten only.
5. Push the light housing back into place.
6. Install the clear plastic shield that covers the sample probe area.
7. Plug the Luminex XYP instrument and the Luminex 200 analyzer into an approved outlet.
Turn on the system components. The first time the software starts up, you have to
acknowledge the license agreement. Read and accept the agreement to continue.
Adjust the Sample Probe Vertical Height
To adjust the sample probe vertical height, perform the following:
In Vitro Diagnostic Use
Luminex 200 System Setup
27