9. Connect the sheath fluid intake line (white fitting) to the Sheath In connector on the front

of the Luminex SD system.

FIGURE 13.

Luminex Sheath Delivery Device

1

Power and Fault indicators

5

Sheath Out connector (blue)

2

Prime button

6

Sheath In connector (white)

3

Regulator adjust

7

Reservoir window

4

Air In connector (green)

10. Lower the stainless steel filter end of the sheath fluid line to the bottom of a full 20 liter

box of Luminex xMAP Sheath Fluid. Secure the cap on the sheath fluid box. Place the
sheath fluid container so that the cap is on the top.

11. Ensure that the sheath fluid box is at least three feet below the Luminex SD system,

otherwise it will overfill.

Luminex

®

200

Installation Manual

30