Assign sample dilution factors, Assign sample dilution factors -64 – Luminex 100 IS User Manual, Version 2.3 User Manual
Page 106

Luminex 100 IS User Manual Version 2.3
x
MAP Technology
5 - 64
PN 89-00002-00-071 Rev. C
To edit a patient list:
1. On the File menu, select Edit Patient List. The Open Batch
dialog box opens.
2. Select the desired batch containing the patient list that you want
to edit and click Select. The Edit Patient List dialog box opens
similar to Figure 5-33.
3. Edit the desired patient and associated dilution factor
information. Click Finish to accept your edits and close the
dialog box.
Figure 5-33. Edit Patient List Dialog Box
Assign Sample
Dilution Factors
Indicate the sample dilution factor so the sample analysis in
quantitative tests is accurate. The system multiplies the result by the
dilution factor for reporting.
Do not use dilution factors for qualitative testing.
You define the dilution factors on a sample-by-sample, or patient-by-
patient basis. The list displays each sample’s accession number and
dilution factor in relation to its well position.
To assign sample dilution factors:
1. On the File menu, click New Batch. The Open Template dialog
box opens.
2. Select the desired template and click Select. The Luminex
Batch Setup dialog box opens. See Figure 5-34.