Figure 5-38. Plate Layout Tab Command Menu
11. Establish preliminary off-plate commands. The system performs
preliminary off-plate commands before the first well command.
To establish a preliminary off-plate command, right-click
anywhere over the plate layout and select Preliminary Off-Plate
Cmd(s) from the menu.The command list dialog box appears
similar to Figure 5-40.
Click the OK button next to the Command List after you select
the desired commands. The white corner marker in the top left
corner of the plate (above the letter A) turns green to indicate an
off-plate command is established. The selected commands run
before the plate begins.
Figure 5-39. Preliminary Off-Plate Command (Corner) Marker
To clear preliminary off-plate commands, right-click on the corner
marker or plate layout and select Clear Off-Plate Cmd(s). See
Figure 5-39.