Luminex 100 IS User Manual, Version 2.3 User Manual
Page 189
Advertising

x
MAP Technology
Luminex 100 IS System Setup
PN 89-00002-00-071 Rev. C
B - 7
3. Unscrew the cap from the Cheminert fitting.
4. Insert the sample probe into the sample probe holder. Insert the
Cheminert fitting into the sample probe holder, turning it
clockwise. Be careful that the threads are correctly aligned. Hand
tighten only. See Figure B-7.
Figure B-7 Insert Sample Probe and Cheminert Fitting
5. Push the light housing back into place.
6. Install the clear plastic shield that covers the sample probe area.
Figure B-8 Sample Probe Area with Clear Plastic Shield in Place
1. Cheminert Fitting
3. Sample Probe
2. Sample Probe Holder
1
3
2
Advertising