Max Machinery 121 INDICATOR User Manual

Page 47

Advertising
background image

121-000-350 © 2001 Max Machinery, Inc.

Pg 47

ENTER

STOP

ENTER

STOP

8.2.8
Calibrate CH2 0mA
Submenu Group

8.2.9
Calibrate CH2 20mA
Submenu Group

Calibrate ch2 0mA
Iin=TB1-8 GND=TB1-4

Calibrate ch2 0mA

0 CALIBRATING ——

Calibrate ch2 0mA

*** DONE ***

Calibrate ch2 0mA
Iin=TB1-8 GND=TB1-4

Calibrate ch2 20mA
Iin=TB1-8 GND=TB1-4

Calibrate ch2 20mA

0 CALIBRATING ——

Calibrate ch2 20mA

*** DONE ***

Calibrate ch2 20mA
Iin=TB1-8 GND=TB1-4

Advance to

Calibrate ch1 0V

To Calibrate: Connect Current Source (+) TB1-8, (-) TB1-4.
Input 0mA and press Enter.

This message is displayed during calibration.

This message is displayed when the 0mA calibration is fin-
ished.

The display will automatically return to the Calibrate CH2
0mA submenu. Press the Down arrow key to advance to the
CH2 20mA calibration.

To Calibrate: Connect Current Source (+) TB1-8, (-) TB1-4.
Input 20mA and press Enter.

This message is displayed during calibration.

This message is displayed when the 20mA calibration is fin-
ished.

The display will automatically return to the Calibrate CH2
20mA submenu. Press the Down arrow key to advance to
the CH1 0V calibration.