Max Machinery 121 INDICATOR User Manual

Page 49

Advertising
background image

121-000-350 © 2001 Max Machinery, Inc.

Pg 49

ENTER

STOP

ENTER

STOP

ENTER

STOP

8.2.12
Calibrate CH2 0V
Submenu Group

8.2.13
Calibrate CH2 10V
Submenu Group

8.2.14
Calibrate 100 ohm
RTD
Submenu Group

Calibrate ch2 0V
Vin=TB1-5 GND=TB1-4

Calibrate ch2 0V

0 CALIBRATING ——

Calibrate ch2 0V

*** DONE ***

Calibrate ch2 0V
Iin=TB1-5 GND=TB1-4

Calibrate ch2 10V
Iin=TB1-5 GND=TB1-4

Calibrate ch2 10V

0 CALIBRATING ——

Calibrate ch2 10V

*** DONE ***

Calibrate ch2 10V
Iin=TB1-5 GND=TB1-4

Advance to

Calibrate 100ohm RTD

Calibrate 100ohm RTD
JMP TB1-6,7 100R=7,8

Calibrate 100ohm RTD

0 CALIBRATING ——

Calibrate 100ohm RTD

*** DONE ***

Calibrate 100ohm RTD
JMP TB1-6,7 100R=7,8

Advance to

Calibrate 4mA out

To Calibrate: Connect Voltage Source (+) TB1-5, (-) TB1-4.
Input 0V and press Enter.

This message is displayed during calibration.

This message is displayed when the 0V calibration is finished.

The display will automatically return to the Calibrate CH2 0V
top-level menu. Press the Down arrow key to advance to the
CH2 10V calibration.

To Calibrate: Connect Voltage Source (+) TB1-5, (-) TB1-4.
Input 10V and press Enter.

This message is displayed during calibration.

This message is displayed when the 10V calibration is fin-
ished.

The display will automatically return to the Calibrate CH2 10V
top-level menu. Press the Down arrow key to advance to the
100 ohm RTD calibration.

To Calibrate: Connect a jumper wire between TB1-6 and TB1-
7, Place a 100 ohm 0.1% resistor between TB1-7 and TB1-8.
Press enter to calibrate.

This message is displayed during calibration.

This message is displayed when the RTD calibration is fin-
ished.

The display will automatically return to the Calibrate 100 ohm
RTD top-level menu. Press the Down arrow key to advance to
the 4mA out calibration.