Multichannel Systems Roboocyte2 Manual User Manual
Page 75

Recording with Roboocyte2 Control Software
71
Installing the Well Plate
Click the alignment icon
in the main window. The "Preparation" window
of the "Alignment Wizard" dialog appears.
Select “Change Plate Pos” to move the carrier to the front of the table. Remove the used well
plate if necessary. Place the well plate which is defined in "Plate Type" in correct orientation
on the carrier.
Place the well plate onto the carrier and press its side against the springs in the lower middle and
lower right corner of the carrier. Then put the rest of the plate down onto the carrier. The well
plate position is correct when there is no play between the well plate bottom and the carrier
surface at all edges and when the well plate has full contact to the left and upper side of the
carrier holding frame.
Important: The well plate should be immovably fixed on the carrier, otherwise the impalement
of the oocytes will not be successful!