4 select the sterilization temperature, 5 running the sterilization program, 5 replacement of the door seal – Nova-Tech B4000-M User Manual

Page 10

Advertising
background image

1 2

7

* Verify all samples are not sealed, capped or closed.

* Do not overload the trays above the stated limit (see Appendix 1).

* Do not stack the trays one above the other or put them in direct contact with

the walls of the sterilization chamber.

* When handling the trays, always use the tray removal tool.

* Wrap each sample separately, if samples are wrapped together verify that

they are of the same material.

* Seal the wrap with sterilization adhesive ribbon or by a thermal sealer.

* Never seal with metallic clips, pins, etc. as this jeopardizes the maintenance

of the sterility;

* When using sterilization paper, set the plastic part downward (tray side)

and the paper side facing upward.

Always wrap samples.

6.4 Select the sterilization temperature

Select the sterilization temperature according to

your sample. See the Appendix 2.

6.5 Running the sterilization program

After selecting temperature, insert the sample into

the chamber using the tray handling tool.

Instructions manual

Instructions manual

8.5 Replacement of the door seal

8.5.1 Fully open the door.

8.5.2 Remove the door seal carefully by pulling it by hand away from

the door

8.5.3 Clean the new door seal carefully with a soft cloth saturated

with distilled water.

8.5.4 Moisten the new seal with medical disinfectant .

8.5.5 Insert the new seal following the instruction below:

1). Press on the top of the seal.

2). Press on the bottom of the

seal.

3). Press on the left and right

portions of the seal.

4). Press the remaining seal surface

and ensure that it is smooth.

Caution: Always ensure the chamber and the door have
been cooled down before attempting to change the seal.