4. Installation

*

*

*

*

*

Ensure that the sterilizer is installed with 2.5in. (10cm) ventilation space

on all sides of the sterilizer, and 5 in. (20cm) on top side. The clearance

required to open the door is 15.5in. (40cm).

The sterilizer should be placed on a level worktable.

Do not cover or block the door, ventilation or radiation openings on the

sterilizer.

Do not install the sterilizer near a sink or in a location where it is likely to

be splashed.

Do not install the sterilizer nearby a heat source.

Instructions manual

Instructions manual

For additional information on any of the error codes listed in section

10, please contact your sales representative, or contact Biomega

Research Products Service Department at 1-908-769-5555. Please

have the unit’s serial number (located on the back panel of the

instrument) available when calling. Do not send in a unit for service

without first calling to obtain a repair authorization number and a

decontamination form. The unit should be properly packed to avoid

damage. Any damage resulting from improper packing shall be the

responsibility of the user.