Draining the water tanks, Maintenance, 1 cleaning the door seal – Nova-Tech B4000-M User Manual
Page 12

1 0
9
7. Draining the water tanks
1.
Connect the supplied drain hose to
the drain valve by pressing it on firmly.
2.
Set the drain valve to the open
position by turning it counter clockwise
3.
Pull the drain valve outward, the
tank will begin to drain through the hose.
4.
After you have completed draining the
water, push the drain valve inward and
set to the closed position
Instructions manual
Instructions manual
7. Maintenance
Frequency
Operation
Daily
Clean the door seal
Wipe dry the inner chamber
Weekly
Clean the clean water tank
Clean the inner-chamber
Replace the door seal
8.2 Cleaning the clean water tank
Clean the clean water tank every
week with medical disinfectant .
8. Maintenance
The sterilizer includes two draining valves. The left valve is used
for draining the used water tank. The right valve is used for draining
the clean water tank. Always be sure that the hose is draining into
a vessel with at least 2L capacity.
Annually
8.1 Cleaning the door seal
Clean the door seal daily with a softy cloth saturated in
distilled water