7 start the sterilization program, 8 sterilization cycle completion, 3 cleaning the inner-chamber, weekly – Nova-Tech B4000-M User Manual

Page 14: 4 door adjustment

Advertising
background image

11

8

6.6

After the samples are loaded, close

and lock the door by turning the door
handle clockwise until it stops.

Caution: You must turn the door handle to the maximum position,

otherwise the machine will alarm and an error message will
be displayed during the cycle.

6.7 Start the sterilization program.

Appendix 2

Press START button - the machine will start
the cycle.

The cycle will take 12-55 minutes.

(See

)

Caution: When you press the "Start" button and the door handle has not

been turned to the maximum position,

will flash

on the Time display window. The cycle will not begin
until the door has been completely closed and the "Start"
button is pressed again.

“do”

After a cycle is completed the time display will show
Once the pressure has reached 0 PSI on the display the door can be
opened and samples can be removed.

“Ed”

6.8 Sterilization cycle completion

Always use the tray handling tool to load or unload the tray in

order to avoid touching the hot surface of the trays.

Instructions manual

Instructions manual

8.3 Cleaning the inner-chamber, weekly

-

Remove the trays and tray rack from the inner-chamber

- Wipe the inner chamber with a soft cloth saturated in distilled water
- Apply the same procedure to the trays and tray rack.

Fig. 2

Fig. 1

Caution:
Never attempt to adjust the door seal when the door is in the
closed position.

8.4 Door adjustment

Below are the instructions for the door adjustment. This should only be
performed if the door is not providing the proper seal or if the door seal has
just been replaced.

8.4.1 Open the door and insert the door adjustment tool in the gap beneath the

plastic cover (Fig. 2).

8.4.2 To tighten the door seal, use this tool to adjust the door adjustment nut

counter clockwise (Fig. 1).

8.4.3 If the door seal is too tight (it is difficult to lock/unlock the door),

you may also adjust the door adjustment nut clockwise to loosen it.

Advertising