Specifications, 14 8. specifications – Nova-Tech pH 700 Ion 700 User Manual

Page 16

Advertising
background image

14

8. SPECIFICATIONS

pH Range

0.00 to 14.00 pH

Resolution

0.01 pH

Accuracy

 0.01 pH + 1 count

Calibration Points

Up to 5 points with Auto-buffer recognition

Buffer Options

USA : pH 1.68, 4.01, 7.00, 10.01,12.45

NIST: pH 1.68, 4.01, 6.86, 9.18 ,12.45

Slope Display

Yes (with offset)

mV Range

 2000 mV

Resolution

0.1 mV (

 199.9 mV), 1 mV beyond  200 mV

Accuracy

 0.2 mV ( 199.9 mV),  2 mV beyond  200 mV

Offset Adjustment

Up to ±150 mV

Temperature Range

0.0 to 100.0 ºC / 32.0 to 212.0 ºF

Resolution

0.1 ºC / ºF

Accuracy

± 0.5ºC / ± 0.9ºF

Compensation

Automatic or Manual (0 to 100 ºC / 32 to 212 ºF)

Calibration

Offset in 0.1 º increments; Offset range:

 5 ºC / 9 ºF

Ion Range

0.01 to 2000 ppm

Resolution (ppm)

0.01 (<1); 0.1 (1.0 to 199.9); 1 (>200)

Accuracy

±0.5% FS (mono-valent) ±1.0% FS (di-valent)

Calibration Points

From 2 to 5 consecutive points;

0.1, 1.0, 10.0, 100.0 or 1000.0 ppm

Slope Range

15 mV/decade to 90 mV/decade

Memory

100 data sets

Inputs

BNC, ATC, Reference (Half-cell)

Power

AC/DC 9V, 6W Adapter (100/240 VAC, 50-60Hz)