3. CALIBRATION

3.1 pH Calibration

For best results, periodic calibration with known accurate standards is
recommended. Calibrate with standards that bracket your intended
measuring range while including a neutral standard (pH 7.00 or 6.86).
For example, if you expect to measure samples from pH 6.2 to pH 9.5,
calibration with 4.01, 7.00, and 10.01 will work well.

The 700 series meters can be calibrated with up to 5 buffers. The
non-volatile memory retains all calibration values upon meter shut down.

The following calibration standards are automatically recognized;

USA buffer group

1.68, 4.01, 7.00, 10.01, 12.45

NIST buffer group

1.68, 4.01, 6.86, 9.18, 12.45

See Section 5.6 to change the buffer group

To eliminate temperature errors associated with the pH electrode, attach
the automatic temperature compensation (ATC) probe for best accuracy.
Without temperature compensation, pH accuracy will worsen as samples
deviate from 25ºC and pH 7.

1) Press

as needed to select pH.

2) Dip the pH and ATC electrodes into pH buffer and press

. The

secondary display will lock on the appropriate buffer value. Provide

stirring for best results. When the READY indicator appears, press

to accept. The primary reading will flash briefly before the secondary
display begins scrolling the remaining available buffers.


3)
Rinse the pH and ATC electrodes then dip into the next pH buffer. The

secondary display will lock on the appropriate buffer value. When the

READY indicator appears, press

to accept. The primary reading

will flash briefly then display the percent efficiency (slope) before the
secondary display begins scrolling the remaining available buffers.