Storing and recalling data – Nova-Tech pH 700 Ion 700 User Manual

Page 9

Advertising
background image

1) Connect the ISE and press

as needed to select ppm. Note: The

primary reading will show “- - - -” without a stored ion calibration. The
secondary value is the corresponding mV reading of the ISE.

2) Dip the ISE into your lowest standard solution value and stir. Press

when the secondary reading is stable.

3) Press

or

to match the primary display to your corresponding

ion calibration value (0.10, 1.0, 10, etc).

4) Press

to accept or

to abort. The primary display will show the

next highest calibration standard value.


5) Rinse the ISE then dip it into the corresponding calibration standard

and stir.

6) Press

to accept or

to cancel. The mV/decade slope value will

be displayed briefly if the calibration is successful. “SLP Err” indicates
that the calibration for the current point was not successful. This occurs
when the slope (mV difference between two consecutive points) is
lower than 15mV/decade or higher than 90mV/decade.

7) Revert to step 3) to calibrate additional points or press

as needed

to return to measurement mode at any time. Note: The meter will
automatically return to measurement mode upon successful completion
of the number of specified calibration points. To change the number of
ion calibration points see Section 5.6.

4. STORING AND RECALLING DATA

The 700 series meters can retain up to 100 points into memory for
later retrieval.

1) In the measurement mode, press

to insert the measured value into

memory. The stored memory location value (StO) is briefly displayed.

7