MEASUREMENT USING TRICHECK

1. Press the

START

button.

All of the display segments are displayed.

Zero (0) is displayed blinking briefly and the

first measurement starts.

2. When the measurement is complete, the

systolic and diastolic pressure readings

and pulse rate are displayed, then a

one-minute countdown starts for the second

measurement.

3. After one minute, the second measurement

starts.

4. When the measurement is complete, the

average readings of the first and second

measurements are displayed, then a

one-minute countdown starts for the third

measurement.

5. After one minute, the third measurement

starts.

6. When the measurement is complete, the

average readings of the three measurements

are displayed and stored in memory.

Notes:

During the measurement, the TriCheck

symbol is displayed.

To cancel the measurement, press the START

button. In this case, no data is stored in

memory.

When the cuff fit error symbol appears during the first measurement, cancel

the measurement, reapply the cuff correctly and start a new measurement.

After measurement, the average data of the three measurements is stored in

memory. No data will be stored, when canceling the measurements, or before

completion of the third measurement.

START

At heart level

Press

All of the display

segments displayed

First measurement

Second measurement

Third measurement

Results of the

first measurement

Average of the first and

second measurements

Average of the

three measurements

One minute after

One minute after