EMC TABLE

Guidance and manufacturer’s declaration – electromagnetic immunity

The A&D unit is intended for use in the electromagnetic environment specified below. The
customer or the user of the A&D unit should assure that it is used in such an environment.

Immunity
test

IEC 60601
test level

Compliance
level

Electromagnetic
environment – guidance

Electrostatic
discharge (ESD)
IEC 61000-4-2

± 6 kV
contact
± 8 kV

air

± 6 kV
contact
± 8 kV

air

Floors should be wood,
concrete or ceramic tile.
If floors are covered with

synthetic material, the
relative humidity should
be at least 30%.

Electrical fast
transient/burst
IEC 61000-4-4

± 2 kV for power
supply lines
± 1 kV
for input/output

lines

± 2 kV for power
supply lines
± 1 kV
for input/output
lines

Mains power quality
should be that of a typical
commercial or hospital
environment.

Surge
IEC 61000-4-5

± 1 kV
line to line
±2 kV
line to earth

±

1 kV

line to line

±

2 kV

line to earth

Mains power quality
should be that of a typical

commercial or hospital
environment.

Voltage dips, short

interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11

< 5% U

T

(> 95% dip in U

T

)

for 0.5 cycle
40% U

T

(60% dip in U

T

)

for 5 cycles
70% U

T

(30% dip in U

T

)

for 25 cycles
< 5% U

T

(> 95% dip in U

T

)

for 5 s

< 5% U

T

(> 95% dip in U

T

)

for 0.5 cycle
40% U

T

(60% dip in U

T

)

for 5 cycles
70% U

T

(30% dip in U

T

)

for 25 cycles
< 5% U

T

(> 95% dip in U

T

)

for 5 s

Mains power quality should

be that of a typical
commercial or hospital
environment. If the user of
the A&D unit requires
continued operation during

power mains interruptions,
it is recommended that the
A&D unit be powered from
an uninterruptible power
supply or a battery.

Power frequency
(50/60 Hz)

magnetic field
IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic
fields should be at levels

characteristic of a typical
location in a typical
commercial or hospital
environment.

NOTE : U

T

is the AC mains voltage prior to application of the test level.