Agilent Technologies Option H48 Multiport Test Set Z5623A User Manual
Page 53
Advertising

Agilent Technologies Z5623A Option H48
4-3
Calibrating the Network Analyzer
Performing the Calibration
Step 2. Press the Preset button on the Network Analyzer.
Step 3. Verify that the Ready light on the ECal Module is green.
Step 4. On the network analyzer, click Calibration on the menu bar, then click
the Calibration Wizard button. See
Figure 4-2
.
Figure 4-2
Calibration Menu on Network Analyzer
Step 5. When the window shown in
Figure 4-3
comes up, click the Cal Type
button. The Cal Type window shown in
Figure 4-4
will then pop up.
Figure 4-3
Cal Wizard Window
The currently selected calibration type is “None”.