Agilent Technologies Option H48 Multiport Test Set Z5623A User Manual
Page 84
Advertising

5 -16
Agilent Technologies Z5623A Option H48
Performance Verification
Setting Up Limit Testing
Step 2. When the Save As window comes up as shown in
Figure 5-15
, type a file
name that includes the words, “insertion loss,” and the date of the
calibration listed in
Table 4-1
on
page 4-13
. For example, if the
calibration file is named, “cal_06_17_01”, an appropriate file name for
this step would be, “insertion_loss_with_06_17_01_cal”. See
Figure
5-16
. After entering the file name, record it in
Table 5-3 on page 5-7
.
Next, click the Save button.
Figure 5-15
“Save As...” Window
Figure 5-16
Specifying The Name Of The Limit Test File For Insertion Loss