Turn on the cal_on_no, Feature – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 136

Advertising
background image

Calibration Procedures Model 9110EH Instruction Manual

Wait until STABIL

value falls below

0.5 ppb. This may

take several

minutes.

ZERO CAL M NOX STB=XXX.X PPB NOX=X.X

< TST TST > ZERO CONC

ZERO CAL M NOX STB=XXX.X PPB NOX=X.X

< TST TST > ENTR CONC

EXIT

If the ZERO button does
not appear, see Chapter

11 for troubleshooting.

Analyzer enters ZERO CAL

mode

EXIT terminates

calibration and returns to

SAMPLE mode.

ZERO CAL M NOX STB=XXX.X PPB NOX=X.X

< TST TST > ZERO CONC

EXIT

ZERO CAL M NOX STB=XXX.X PPB NOX=X.X X

< TST TST > CAL CALZ CALS

SETUP

SAMPLE RANGE = 500.0 PPB NOX= X.X

< TST TST > CAL CALZ CALS

SETUP

SAMPLE NOX STB=XXX.X PPB NO = X.X

< TST TST > CAL CALZ CALS

SETUP

Set the Display to show the

NOX STB test function.

Pressing ENTR changes the
calibration of the instrument.

Turn ON the CAL_ON_NO

2

feature:

SAMPLE RANGE = 500.0 PPB NOX=X.X

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS RNGE PASS CLK MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM VARS DIAG

EXIT

SETUP X.X

0 ) DAS_HOLD_OFF=15.0 Minutes

NEXT JUMP

EDIT PRNT EXIT

SETUP X.X

ENTER VARS PASS: 818

8 1 8

ENTR EXIT

EXIT ignores the new setting.

ENTR accepts the new setting.

SETUP X.X

8) CAL_ON_NO

2

= OFF

PREV NEXT JUMP

EDIT PRNT EXIT

SETUP X.X

8) CAL_ON_NO

2

= ON


ON

ENTR EXIT

Toggle this key to change setting

Press NEXT until …

122 M9110EH Rev 0