Figure 3-6, D figure 3-6 illu, Model 9110eh – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual
Page 30: Getting started model 9110eh instruction manual, Insufficient gas flow can build up no

Getting Started Model 9110EH Instruction Manual
External Zero
Air Scrubber
Filter
Pump
MODEL 701
Zero Air
Generator
Source of
SAMPLE Gas
VENT if input is pressurized
Zero/Span Valves – Option 50
MODEL 700
Gas Dilution Calibrator
(with O
3
generator option)
VENT
MODEL
9110EH
Sample
Zero Air
Exhaust
Span
Source of
SAMPLE Gas
VENT if input is pressurized
Internal Zero/Span Option (IZS) – Option 51
Pump
MODEL
9110EH
Sample
Zero Air
Exhaust
Span
External Zero
Air Scrubber
Particulate
Filter
Calibrated
NO gas
(high
concentration)
Figure 3-6: Basic Pneumatic Connections for Units With Valve Options
Once the appropriate pneumatic connections have been made, check all pneumatic fittings
for leaks using the procedures defined in Section 11.5.
WARNING
Gas flow must be maintained at all times for units with IZS Options installed. The
IZS option includes a permeation tube which emits NO
2
. Insufficient gas flow can
build up NO
2
to levels that will damage the instrument. Remove the permeation
device when taking the analyzer out of operation.
16 M9110EH Rev 0